Microscope Application Program http://www.ViTiny.com
Copyright ©2008-2012 MicroLinks Technology Corp.
Fig.5-47 Setting scale dialog
Scale default file
Scale default files have two types “Default Height Magnification” and Default
Height Magnification” shown in Fig.5-48
Fig.5-48Scale default File
When the microscope is observing object with contacted distance, it has two focus
points. Thus, it will have two kinds of scale value which are for low and high mag.
When in low mag, the horizontal and vertical scale is 2.24 mm and 1.68mm. And
when in high mag, the horizontal and vertical scale is 0.56mm and 0.42mm. If the
default value is not right for the user, the scale value can reset or create a file for new
value by calibrate the horizontal and vertical scale according to the user‟s preferences.
If want to return the default value, chick “Scale default” box than it will delete the
new add file and return to the two default setting.